Medela Invia Liberty Manual De Uso página 19

Ocultar thumbs Ver también para Invia Liberty:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3. Click in Invia Liberty
Canister
Open external packaging.
4. Switch Invia Liberty
Pump ON
by pressing [
].
5. Conduct functional check
to ensure adequate seal.
6. If functional check was successful, switch OFF Invia Liberty Pump by pressing
[
> 3 seconds ]. If the self-test is not successful, follow instructions shown on the
display or see the Alarms chapter.
7. Connect the Invia Liberty Tubing Connector to the wound drain tubing or the tubing of the
External Suction Interface (ESI). For dressing applications, please refer to Instruction for
Use provided with the Invia and Avance Dressing Kits. Switch Invia Liberty pump on as
described in the User Modes.
1.
2.
1.
1.
CAUTION
Ensure tubing is not bent during the functional check.
1. Slip the pegs at the bottom
of the cannister into the slots
at the base of the pump.
Lift cannister towards pump
until it clicks into place.
2. "Click" into place.
Canister secures tubing
1. The pump will do self-test
and start running imme diately.
You should hear the motor
running for a short time until
the pressure is built up. If the
motor keeps running for more
than 30 sec. – check system
and retry.
1. Seal off the conical
connector of the tubing with
your thumb (through the
inner packaging).
19