Troubleshooting; Corrective Action - SATA adam 2 Serie Instrucciones De Servicio

Ocultar thumbs Ver también para adam 2 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 88
EN
Operating instructions pressure measuring devices SATA adam 2, SATA adam 2 U

10. Troubleshooting

Malfunction
No pressure is
displayed.
Display cannot
be removed at
all or only with
difficult .
Air leaks be-
tween display
and connection
module dock
Air leaks from
the ball valve [1-
8] (when using
without display)
Display does not
snap-in
Digital display is
either black or
blurred.
82
Cause
• Display has not prop-
erly snapped into the
dock
• Battery is empty
• Regulation wing is not
in horizontal centre
position
• Slide-in groove is
clogged.
• Locking clip has not
snapped in.
• Special O-ring [1-6] is
swollen due to soaking
in solvent.
O-Ring beneath the
locking clip is missing or
damaged.
Ball valve is clogged
• Locking clip is dam-
aged.
• Slide-in groove of the
connection module is
clogged or damaged
Display has been exces-
sively exposed to clean-
ing solution.

Corrective action

• Slide in the display
until it snaps home,
chapter 7.1
• Replace the battery
• Place regulation wing
in horizontal centre
position, chapter 7.1
• Clean the slide-in
groove.
• Press in the locking
clip, chapter 7.1
• Avoid soaking in sol-
vent, chapter 8
Replace the O-ring,
chapter 9
Clean ball valve, chap-
ter 8
• Replace locking clip,
chapter 9
• Clean the groove
or replace the dock,
respectively
Cleaning, chapter 8,
replacement of the dig-
ital unit

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Adam 2 miniAdam 2 displayAdam 2 u

Tabla de contenido