Location Name
•
This selection allows the customization of a Test Stand location
name by the user.
•
This name can be up to eight characters and will be used to
tag information provided to the network interface and memory
card.
•
Use
,
, and ENTER ( ) to select each individual character of
the location name.
•
All eight characters must be entered (use spaces for unwanted
characters if needed).
NOTE: The following options are only available on "MASTER" Test
Stands (Test Stands that do not have another Test Stand con-
nected on the right).
Print Receipts
•
Use
,
, and ENTER ( ) to select ON or OFF to enable or
disable print receipts.
•
A test receipt will print when this option is turned ON.
Email Overdue Now
•
This selection immediately sends out an email message for each
overdue instrument located in the internal database.
•
Before using this option, make sure to properly set up email alerts
in the Network Interface (requires an email server –SMTP- and
valid email address).
Synch All
•
This selection will copy the settings from the "MASTER" (rightmost)
Test Stand to all other Test Stands.
•
Use of this command changes cylinder setup and may take up to
one minute.
•
Make sure the "MASTER" Test Stand is set up with all
cylinders used in the setup and not just the ones needed by
itself.
Network Reset
•
This selection will reset the network interface back to factory
default settings.
3-7