Instrument alarm

Intermittent alarm:
Special symbol »
« displayed on the right side of the display:
The instrument is not ready for operation.
Commission maintenance personnel or the Dräger Service Centre to
eliminate the error.

Activating the Info mode

In measuring mode, press the [OK] key for approx. 3 seconds.
If any warning or fault messages exist, the corresponding note or error codes
are displayed (see Technical Handbook). Press the [OK] key successively
for the next display. The peak values and the exposure values TWA and
STEV will be displayed.
If no key is pressed for 10 seconds, the instrument returns automatically to
measuring mode.

Calling the Info-Off mode

With the instrument switched off, press the [+] key for approx. 2 seconds.
The name of the gas, measuring unit, and measuring range limit value are
displayed for all channels.
Pressing the [+] key again exits the Info Off Mode (or via timeout).

Calling the Quick Menu

In measuring mode, press the [+] key three times.
If functions in the quick menu are activated using the PC software "Dräger
CC Vision", you can select these functions using the [+] key. If no functions
have been activated in the quick menu, the instrument remains in measuring
mode.
Possible functions:
1. Bump test
2. Fresh-gas adjustment
3. Delete peak values
Press the [OK] key to call the selected function.
Press the [+] key to cancel the active function and to switch to measuring
mode.
If no key is pressed for 60 seconds, the instrument returns automatically to
measuring mode.
1)
The fresh-gas adjustment / zero-point adjustment is not supported by the DrägerSensor DUAL
IR CO
, DrägerSensor IR CO
or the DrägerSensor XXS O
2
2
sensors can be made with the Dräger CC Vision PC software. A suitable zero gas free from
carbon dioxide and ozone (e.g. N
) must be used.
2
1)
. A zero-point adjustment of these
3
23