How To Use - Topcom 200 Manual De Usuario

Ocultar thumbs Ver también para 200:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 19
8

HOW TO USE

• Please make sure to use a new and clean probe cover to ensure accurate measurement.
• Press the Start - button (1) to turn the power on. The display will shows all symbols.
Before inserting the probe:
- For children < 1 year : Pull the ear down and back.
- For children > 1 year to adult : Pull the ear up and back.
• While holding the ear, insert the probe to seal the ear canal, then press the Start - button
(1) once. You will hear a short beep.
• When you hear a short melody, the measurement is finished.Take out the thermometer
from the ear canal. The measured temperature is displayed. The result will be stored auto-
matically into the memory. If the memory is full (30 records), the oldest measurement result
will be removed.
It is recommended that you take three temperatures in the same ear under the following
circumstances. If they are different, use the highest reading.
• Infants in the first 3 months of life.
• Children under three years old who have a condition such as acompromised
immune system and for whom the presence or absence of fever is critical.
• When you are first learning to use the ear thermometer until you are comfortable
with the technique and are gaining consistent readings
Important: As with any type of thermometer, slight temperature variations (+/- 0.2 -
0.3°C) can happen, if consecutive measurements are taken.
9
DISPLAY ERROR MESSAGES
'HI'
The measured temperature is higher than 42,9 °C.
'LO'
The measured temperature lower than 32 °C
'ErrE'
The thermometer temperature is not within the operation specifications (<10°C or >40°C)
'ErrU'
Don't measure after pressing 'Measure'. Follow the steps mentioned in the user guide.
'ErrH'
An error is detected during self-check. Remove the batteries for 10 seconds and try again.
'
'
Low battery. Replace the battery.
10
TIME SETTINGS
• Power on the unit by pressing the Start - button (1) once. You will hear a short beep. Time mode is displayed.
• Press and hold the Memo - button (4). The hour digits are blinking.
• Press the Memo - button (4) repeatedly to increase the digit.
• Press the Talk - button (3) to confirm. The minutes digits are blinking.
• Repeat previous steps for setting the minutes, year, month and day.
• After setting the day, the unit will enter time mode.
11
TALK MODE
The result of each measurement is broadcasted through the speaker.
To broadcast the time, press the Talk - button (3) in Time mode.
• To disable the talk mode, press and hold the Talk - button (3) for 2 seconds. You hear one beep tone.
• To enable the talk mode, press and hold the Talk - button (3) again for 2 seconds. You hear a double beep
tone.
12
MEMORY
The Ear Thermometer will automatically store the last 30 measurements.
• Power on the unit by press the Start - button (1) once. You will hear a short beep. Time mode is displayed.
• Press the Memo - button (4). The last result is displayed.
• Press the Memo - button (4) repeatly, the display will show the temperature of the previous measurements.
• Press and hold the Memo - button (4) repeatly to scroll rapidly.
TOPCOM EAR THERMOMETER 200
5

Publicidad

Tabla de contenido
loading

Tabla de contenido