Intended Use; Service Life - Ottobock 2R40 Serie Instrucciones De Uso

Ocultar thumbs Ver también para 2R40 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 23
other manufacturers that have compatible modular connectors has not
been tested.
2R41=1: Only combine the product with foot adapter 2R40=1.
2R41=2: Only combine the product with foot adapter 2R40=2.

2 Intended use

2.1 Indications for use
The product is intended exclusively for lower limb exoprosthetic fit­
tings.
2.2 Area of application
The product was developed for use in a children's prosthesis.
The maximum approved body weight is specified in the technical
data (see Page 13).
2.3 Environmental conditions
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative
humidity 20 % to 90 %, no mechanical vibrations or impacts
Allowable environmental conditions
Temperature range: –10 °C to +45 °C
Moisture: relative humidity: 20% to 90%, non-condensing
Unacceptable environmental conditions
Chemicals/liquids: fresh water, salt water, perspiration, urine,
acids, soapsuds, chlorine water
Solids: dust, sand, highly hygroscopic particles (e. g. talcum)

2.4 Service life

These prosthetic components were tested by the manufacturer
through 3 million load cycles. This would correspond to a maximum
service life of 5 years for an adult's prosthesis. A general service life
cannot be determined for these prosthetic components when used as
part of a child's prosthesis since the strain varies widely depending
on the intensity of use.
10
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or
CAUTION
injury.
3.2 General safety instructions
CAUTION
Excessive strain on the product
Risk of injury due to breakage of load-bearing components
► Use the product according to the specified area of application
(see Page 10).
CAUTION
Unallowable combination of prosthetic components
Risk of injury due to breakage or deformation of the product
► Only combine the product with prosthetic components that are
approved for that purpose.
► Based on the instructions for use of the prosthetic components,
verify that they may be combined with each other.
CAUTION
Use under unallowable environmental conditions
Risk of injury due to damage to the product
► Do not expose the product to unallowable environmental condi­
tions.
► If the product has been exposed to unallowable environmental
conditions, check it for damage.
► If damage is apparent or in case of doubt, do not continue using
the product.
► Take suitable measures if required (e.g. cleaning, repair,
replacement, inspection by the manufacturer or a specialist
workshop etc.).

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

2r41 serie2r484r66

Tabla de contenido