Linearity Calibration - OHAUS Defender 5000 Manual Del Usuario

Ocultar thumbs Ver también para Defender 5000:
Tabla de contenido

Publicidad

Defender 5000 Indicators
A suggestive message shows on the screen.
The message 'Completed' will be displayed on the
screen.
Exit Span Calibration by pressing the Softkey
corresponding to the icon
To return to the Main Menu, press the Softkey
corresponding to the icon
Note: Span Calibration should be performed after
Zero Calibration.

3.3.3 Linearity Calibration

Linearity calibration uses 3 calibration points. The full calibration point is established with a weight
on the scale. The mid calibration point is established with a weight equal to half of the full calibration
weight on the scale. The zero calibration point is established with no weight on the scale. The full
calibration and mid calibration points can be altered by the user during the calibration procedure.
Calibration procedures:
Long press the button
Menu.
Press the Softkey corresponding to the icon
enter the Calibration sub-menu.
Scroll to Linearity Calibration using the Softkey
corresponding to the icon
Press the Softkey corresponding to the icon
initiate Linearity Calibration.
Clear the pan and press the Softkey corresponding to
the icon
.
Put the calibration mass of the specified weight on
the pan, and then press the Softkey corresponding
to the icon
for confirmation. To change to a
different calibration point, input the value desired,
and then place the corresponding weight on the pan
for calibration.
.
.
to enter the Main
to
.
to
EN-19

Hide quick links:

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Td52pTd52xwDefender td52xwDefender td52p

Tabla de contenido