Press the [OK] key to start the channel selection.
In the case of the CO
channel there is a 2-stage calibration routine:
2
The zero calibration takes place first and is then followed by the span calibration.
–
The display shows the gas of the first measuring channel, e.g. » CH
Press the [OK] key to start the adjustment function of this measuring channel,
or use the [ + ] key to select another measuring channel (O
CO - ppm, etc.).
–
The test gas concentration is displayed.
Press the [ OK ] key to confirm the test gas concentration or use the [ + ] key
to change the test gas concentration and complete the process by pressing
the [ OK ] key.
–
The measurement value flashes.
Open the test gas cylinder valve to let gas flow over the sensor with a
volume flow of 0.5 l/min.
–
The displayed, flashing measurement value changes to the value according
to the supplied test gas.
When the displayed measurement value is stable (after at least 120 seconds):
Press the [ OK ] key to perform the adjustment.
–
The display containing the current gas concentration changes with the
display » OK «.
Press the [ OK ] key or wait for approx. 5 seconds to end the adjustment of
this measuring channel.
–
The next measuring channel is displayed for adjustment if necessary.
–
After the adjustment of the last measuring channel, the instrument changes
to the measuring mode.
Close the test gas cylinder valve and remove the instrument from the
calibration cradle.
If a fault has occurred during the span calibration:
–
The fault message »
« appears and »
respective sensor instead of the measured value.
In this case, repeat the calibration.
Change the sensor if necessary.
12
NOTICE
- Vol.-%, H
2
2
« is displayed for the
Cleaning
The instrument does not need any special care.
Dirt and deposits can be removed from the instrument by washing it with
cold water. A sponge can be used for wiping if necessary.
LEL «.
4
Abrasive cleaning tools (brushes etc.), cleaning agents and cleaning solvents
S - ppm,
can destroy the dust and water filters.
Carefully dry the instrument with a cloth.
Maintenance
The instrument should be inspected and serviced once a year by suitably
qualified persons. Comparisons:
EN 60079-29-2 – Gas detectors - Selection, installation, use and
maintenance of detectors for flammable gases and oxygen
EN 45544-4 – Electrical apparatus used for the direct detection and direct
concentration measurement of toxic gases and vapours - Part 4: Guide for
selection, installation, use and maintenance
National regulations
Recommended calibration interval for measuring channels O
NO
and CO: 6 months.
2
Recommended calibration intervals for the measuring channel IR Ex/CO
12 months.
Calibration intervals of other gases: see Instructions for Use of the respective
DrägerSensors.
See the Technical Manual for details of spare parts.
Storage
Dräger recommends storing the instrument in the charger module
(order no. 83 18 639).
Dräger recommends checking the charge of the power supply at least every
three weeks if the instrument is not stored in the charger module.
CAUTION
, H
S, H
, SO
2
2
2
,
2
:
2