Ottobock 1E91 Runner Instrucciones De Uso página 19

Ocultar thumbs Ver también para 1E91 Runner:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 43
TT bench alignment
Required materials: 662M4 goniometer, 743A80 50:50 gauge, alignment
apparatus (e.g. 743A200 PROS.A. Assembly)
Perform the assembly and alignment of the prosthetic components in the
alignment apparatus according to the following specifications:
   
Sagittal alignment prosthetic foot
a–p position of the reference point to the alignment reference
line: See illustrations at the start of the document
   
Frontal alignment of prosthetic foot
Lateral rotation: 0° (see Page 20)
   
Sagittal alignment pivot point knee joint
a–p position of pivot point to alignment reference line:
-15 mm
   
Connect the prosthetic foot and prosthetic socket using the chosen
adapters. Be sure to follow the adapter instructions for use.
   
Sagittal alignment prosthetic socket
In proximal area centred on alignment reference line
Socket flexion: Individual residual limb flexion + 5°
   
Observe the abduction or adduction position.
TF bench alignment
Required materials: 662M4 goniometer, 743A80 50:50 gauge, alignment
apparatus (e.g. 743A200 PROS.A. Assembly)
Perform the assembly and alignment of the prosthetic components in the
alignment apparatus according to the following specifications:
   
Sagittal alignment prosthetic foot
a–p position of the reference point to the alignment reference
line: See illustrations at the start of the document
   
Frontal alignment of prosthetic foot
Lateral rotation: Specified by the prosthetic knee joint (see
Page 20)
   
Sagittal alignment prosthetic knee joint
a–p position of the reference point to the alignment reference
line: -15 mm
   
Frontal alignment prosthetic knee joint
Lateral rotation: approx. 5°
Bench alignment process
Bench alignment process
Ottobock | 19

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

1e93 runner junior

Tabla de contenido