ON/OFF
START/STOP
The "hold" mode is used to temporarily stop the flow of feed:
• to change the feeding program (installed parameters),
• to change the feed container,
• to administer medication without switching the pump off,
• to silence an alarm and correct problems.
After 3 minutes a two tone audible alarm sounds and the message "PUSH STRT" appears in the display.
Press "START/STOP" to stop the alarm and to extend the hold mode by a further 3 minutes or press "START/
STOP" twice (3 beeps will be heard) to resume programmed settings.
ON/OFF
FILL SET
Ensure the feeding set is not connected to the patients feeding tube when the pump is in the hold mode:
Press and hold the "FILL SET" key for 2 seconds to activate the "FILL SET" maneuver. Release the button as
soon as the pump generates a beep and starts pumping at a flow rate of approx. 700ml/h. During this maneu-
ver the message "FILL SET" appears in the display. The pump will automatically stop when the Flocare
Infinity
™
pump set is completely filled with feed. However this "FILL SET" maneuver can at anytime be stopped
by pushing the "FILL SET" key a second time. The pump will return to the hold mode when the "FILL SET"
maneuver is complete or stopped.
The pump has a liquid crystal display (L.C.D.) with large alphanumeric characters, smaller symbols, words
and a back light.
ZMNL900616_AW.indd 18
SWITCHING "OFF"
Press the "ON/OFF" key and keep it pressed for 2 seconds. A continuous alarm will be
heard and the pump switches off.
The feeding program (installed parameters) and total volume administered since the
last clearance will be retained in the pump's memory.
If the pump shuts down due to low battery voltage, the memory will be retained for
24 hours.
"HOLD" MODE
To temporarily pause the pump, or switch into "hold" mode whilst operating, press
the "START/STOP" key once. Three beeps are heard and the run symbol disappears.
The programmed flow rate (ml/h), volume (DOSE=VOL) and the administered volume
(ml) are retained.
FILLING THE PUMP SET
The FILL SET function offered by the Infinity pump is used to fill a (new) feeding set
with nutrition (or water as the case may be). When the FILL SET function runs all alarms
are deactivated helping avoid alarms being inadvertently triggered (e.g. air alarm).
The FILL SET function should only be used to fill an empty (air-filled) feeding set. Use of the FILL
SET function at any other time may cause the pump to incorrectly calibrate.
DISPLAY
UK-IE
-18-
®