Set Up Mode - Nutricia Flocare Infinity Instrucciones De Uso

Flocare infinity
Ocultar thumbs Ver también para Flocare Infinity:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Condition
OCC OUT
• Calibration not yet completed
(repeated)
LOCK
• Only the current feeding
ER01 - ER99
• The self test detected an electronic
ER40
• Keypad pressed or stuck during
No plug symbol visi-
• The wall outlet doesn't work.
ble, while the pump
• The adaptor is damaged.
is connected to the
mains.

SET UP MODE

The set up mode is used • to set the alarm level,
This set up mode may only be entered by healthcare professionals and authorised personnel trained to use this applica-
tion of the pump.
The patient or his/her relatives without permission of the physician, dietician, nurse or other licensed practitioner, may
not change the settings of the set up mode.
+
ON/OFF
&
2 sec
2 sec
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode.
First the alarm level can be set:
"BEEP HIGH" (+) or "BEEP LOW" (-)
Use the "+" or "-" key to change the setting from "high" to "low" or from "low" to "high".
ZMNL900616_AW.indd 24
Cause
on the current feeding set
schedule is allowed for this patient.
Another feeding program is not
allowed by your healthcare profes-
sional.
error
pump start up.
• to lock the keypad,
• to switch the audible alarm off when a dose is administered,
• to switch the light of the LCD permanently on when connected to the mains,
• to enable connection with a Nurse call or Patient Data
Monitoring System (PDMS).
TO SET THE ALARM LEVEL
+
Beep high
d
-
Beep low
UK-IE
Correction
• Stop the alarm by pressing the "START/STOP" key.
• Start the pump by pressing the "START/STOP" key and
let it run only briefly.
• Stop the pump by pressing the "START/STOP" key insu-
ring that there has been no occlusion out alarm.
• Remove the feeding set from the pump and reinsert the
feeding set in the pump.
• Restart the pump by pressing the "START/STOP" key.
• The programming feature is blocked in the set up mode
of the pump. Ask your healthcare professional to modify
this setting.
• Turn the pump "off", make sure the pump door is closed
and switch the pump back "on". If the error persists,
contact your healthcare provider or refer the pump to
service.
• Restart pump by ensuring keypad is not pressed.
• Connect the pump to another wall outlet.
• Contact your healthcare professional or
Nutricia subsidiary to replace the adaptor.