y
-
Condition
NO SET
d
.
r
d
r
PUSH STRT
END OF DOSE
.
-
e
PROG
.
s
DOOR
AIR
d
n
y
BATT
the pump
remains working
Battery, "E" and "F"
flashing
l
FILL SET
OCC IN
OCC OUT
ZMNL900616_AW.indd 23
UK-IE
Always follow the instructions below in case a problem occurs.
Using methods other than those described may cause the pump to function incorrectly.
Cause
• The set is not fitted or wrongly fitted
in the pump.
• The pressure sensor area is dirty (
Fig 1, position 5 & 6)
• The pump has been untouched in
hold mode for 3 minutes or more.
• The pump administered the installed
dose = volume.
• No flow rate is installed.
Flow rate = 0 ml/h
• The door is not correctly closed.
• The pump detected air in the pump
set.
• The air sensor area is dirty( Fig 1,
position 8).
• The feeding set is not inserted cor-
rectly.
• The battery capacity is too low.
The pump is not able to deliver
highly viscous fluids at high flow
rates with the present charge level
of the battery.
• Battery failure.
• The pump is priming the set.
• The pump detected an upstream
occlusion between the pump and
the feeding bag.
• The pressure sensor area is dirty (
Fig 1, position 5 & 6).
• The pump detected a
downstream occlusion between the
pump and the patient.
• The pressure sensor area is dirty
(Fig 1, position 5 & 6).
PROBLEM SOLVER CHART
Correction
• Stop the alarm by pressing "START/STOP"
• Insert the Flocare
indicated on the blister packaging (or refer to Fig 5a - 5d)
and close the door.
• Restart the pump.
• Clean the sensors, reinsert the feeding set in the pump
and restart the pump.
• Stop the alarm and prolong the hold mode with another 3
minutes by pressing "START/STOP" key.
• Program the pump and start it by pressing the "START/
STOP".
• The programmed dose/volume has been administered
• Press start/stop to switch off the alarm
• Press INFO and then CLR to reset the administered
volume (to reset the flow rate and dose as well, press the
CLR button for 3 seconds)
• Make sure the correct flow rate is programmed.
• Make sure the door is properly closed prior to
starting a feeding program.
• Make sure there is no excessive foam in the reservoir and
ensure there is enough feed in the reservoir for the set
therapy.
• If the feeding reservoir is empty, replace the reservoir and
continue feeding. If necessary, prime the set (FILL SET).
• Make sure the air sensor is clean.
• Make sure the feeding set is properly inserted in the
pump. (Fig 5a - 5d)
• Connect the adaptor to the pump and mains and charge
the pump for approx. 6 hours. During charging the pump
can be used.
• Turn the pump "off", contact your healthcare provider or
refer the pump to service.
• Press the "FILL SET" key another time to stop the pump
and bring it back in the hold mode.
• Stop the alarm by pressing the "START/STOP" key.
• Remove the feeding set out of the pump and check the
permeability by flushing the line.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the pump
and restart the pump.
• Stop the alarm by pressing the "START/STOP" key.
• Remove the feeding set out of the pump and check the
permeability by flushing the line.
• Check the permeability of the feeding tube.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the pump
and restart the pump.
-23-
®