Warnings And Cautions; Alarm Functions And Safety Features - Nutricia Flocare Infinity Instrucciones De Uso

Flocare infinity
Ocultar thumbs Ver también para Flocare Infinity:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

• On a regular basis thoroughly clean all surfaces of the pump (including the sensors and rotor) with warm soapy
water, a 5% bleach solution in water, a multipurpose disinfectant cleaner or a soft cloth.
• The Flocare
Infinity
®
merge the pump!
• Always maintain the rollers on the rotor in a clean state to ensure they spin freely.
• The adaptor normally does not require cleaning. When desired, a dry or slightly damp cloth may be used
to clean the outside surface of the adaptor. Make sure the adaptor is disconnected from the wall outlet.
• Pump disposal: At the end of their service lifetime , the pump and its electrical accessories (AC Adapter
Charger and Data Download Accessory Cable), should be disposed of according to local standards and
regulations governing the disposal of electronic waste (e-waste). Other accessories can be disposed of or
recycled as standard non-hazardous waste.
• The manufacturer recommends an inspection of the pump at an authorised service centre every 2 years.
• If any fault occurs during use, or if the pump is dropped, it should be checked by authorised technical per-
sonnel prior to use.
• Replace the Flocare
growth of harmful bacteria. Dispose of Flocare
• Do not use the Infinity alarm systems to trigger actions related to secondary (electrical) medical devices
(e.g. a volumetric or syringe pump).
• Do not use pump functions (e.g. the fill set function) for any other purpose than described in this manual,
as this may cause the pump to incorrectly calibrate.
• No modification of this equipment is allowed.
• The AIR alarm should not be used to indicate an end of dose. NOTE: do not mute the audio alarm for end
of dose alarm in these instances ('end of dose' will appear in the pump screen).
• Non-liquid food ingredients and mixed or reconstituted foods may have a tendency to occlude flow within
the feeding tube and feeding set. They may also have the tendency to adhere to the tubing wall and
obstruct the enteral feeding pump optical air-in-line sensor. Clinicians and caregivers should carefully
monitor the pump during use with these type of feedings.

ALARM FUNCTIONS AND SAFETY FEATURES

In case any of the problems listed on the problem solver chart occur, the pump delivers an audible and visual
alarm and stops working. The back light of the LCD automatically switches on.
Exception to this is the low battery alarm "BATT", in this situation the pump continues working.
Action in the event of an alarm:
Check the type of alarm displayed by the large display characters.
• Press the "START/STOP" key to stop the audible and visual alarm.
• Correct the cause of the alarm as described in the table.
• Start the pump again by pressing "START/STOP".
The problem solver chart gives a clear explanation of the alarms.
ZMNL900616_AW.indd 22
pump may be rinsed by holding under a stream of warm, clean water. Do not sub-

WARNINGS AND CAUTIONS

Infinity
pump set every 24 hours to maintain delivery accuracy and prevent the
®
®
Infinity
disposable sets properly, as required by local law.
-22-
UK-IE
10/11/16 16:07

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

35679

Tabla de contenido