Entry of formulas. Rules for formula entry, see page 34.
Example: "H2O*C01/C00"
In addition to the formula, a text for result output, the number of decimal places and
a unit for the result output can be selected. "No unit" is selected with the blank string.
In place of "RSX", a result name may be entered (.TextRS). This name is outputted in
the result report, scalc full and scalc srt. It is used for the result and the correspond-
ing mean value.
The limit control for results can also be activated. If a result is out of limit, a message
appears in the result report, E196 is sent, and output line L13 can be set.
4.2.2.23. Mode.Def.SiloCalc.Assign.C24
.Assign.C2X: Assignment to store results in the silo as C2X.
.MatchId:
4.2.2.24. Mode.Def.ComVar.C30
Assignment of common variables.
The values of the common variables are to be found in &Config.ComVar. They can
be viewed and entered there, see 4.2.2.54.
Definition of the report sequence, which is outputted automatically at the end of the
determination. Entries of more than one block have to be separated with ";".
.Internal:
.Assign1: Output to COM1 of the Coulometer.
.Assign2: Output to COM2 of the Coulometer.
4.2.2.26. Mode.Def.Mean.1.Assign
Assignment of the statistics calculations. Valid assignments are a requirement for
statistics calculations. In addition, the statistics calculation must be switched on, see
4.2.2.13. Rules for statistics calculations see page 37.
756/831 KF Coulometer, Instructions of Use
Mode.Def.Formulas.1.Decimal
Mode.Def.Formulas.1.Unit
Mode.Def.Formulas.1.Limits
Mode.Def.Formulas.1.LoLim
Mode.Def.Formulas.1.UpLim
Mode.Def.Formulas.1.Output
Mode.Def.Formulas.2.Formula
etc. up to .9
Mode.Def.SiloCalc.Assign.C25
Mode.Def.SiloCalc.MatchId
Indication which sample identification(s) have to match so that the
results can be combined.
Mode.Def.ComVar.C31
etc., up to .C39
4.2.2.25. Mode.Def.Report.Internal (only 756)
meas crv, comb, mplist, param, calc, scalc full, scalc srt, ff
Mode.Def.Report.Assign1
Mode.Def.Report.Assign2
Internal printer of the Coulometer. (only 756)
Mode.Def.Mean.2.Assign
etc., up to .9
4.2 Remote control commands
0...2...5
6 ASCII characters
ON, OFF
0...±999 999
0...±999 999
active, pulse, OFF
RSX, H2O, CXX
RSX, H2O, CXX
id1, id1&2, all, OFF
RSX, MNX, H2O, CXX
result, water crv, rate crv,
ditto
ditto
RS1, RSX, H2O, CXX
77