.Method: Method used to process the sample, from the method memory or from the
.Id:
.UnitSmpl: If "no unit" is desired for the sample size, the blank string must be en-
.C24, .C25: Results which have been assigned to C24 and C25.
.Mark:
4.2.2.58. SmplData.ONSilo.DelLine
Deletion of a silo line. The line # is deleted with &SmplData.ONSilo.DelLine $G. If a
formerly deleted line is edited again, it becomes valid (function "undelete").
4.2.2.59. SmplData.ONSilo.DelAll
Deletes the entire silo memory. Must be triggered with $G.
4.2.2.60. SmplData.ONSilo.CycleLines
Silo data cycling.
With "ON", executed lines are copied to the next free silo lines, see page 49. Exercise
caution if you edit the silo memory during the determinations!
4.2.2.61. SmplData.ONSilo.SaveLines
Silo lines are not deleted when they are worked off. Assigned results are stored as
C24 and C25. "Save lines" can only be set to "ON" if the silo is completely empty.
Delete the silo, see 4.2.2.60.
4.2.2.62. HotKey.User.Name
Management of user names.
.Name:
.Delete,Name: Deletes selected user name with &HotKey.User.Delete $G.
.List:
4.2.2.63. Info.Report
$G sends the selected report to the COM which is set in
&Config.PeriphUnit.RepToComport:
756/831 KF Coulometer, Instructions of Use
card.
The identifications Id1...Id3 can also be used as sample-specific calcula-
tion constants C21...C23 in formulas.
tered.
Mark of the silo line: "*"=deleted line, "+"=line which is worked off, "-
"= line which is worked off and not valid for silo calculations (deleted),
"/" last worked-off line, where recalculation can still be done. Silo lines
which have been worked off are "read only".
SmplData.ONSilo.DelLine.LineNum
HotKey.User.Delete
HotKey.User.Delete.Name
HotKey.User.DelAll
HotKey.User.List.1.Name
Input of user names.
List of all user names.
Info.Report.Select
result, water crv, rate crv, meas crv, comb,
mplist, param, calc, C-fmla, def, statistics, smpl data,
silo, scalc full, scalc srt, config, user method, all, ff
4.2 Remote control commands
$G
1...255, OFF
$G
ON, OFF
ON, OFF
up to 10 ASCII characters
$G
up to 10 ASCII characters
$G
read only
$G
83