GoodKnight 425 / GoodKnight 425 ST
15.0
Positive Inspiratory Pressure IPAP level set of your
cmH
O
2
device.
You can increase or decrease the positive inspiratory
pressure level of the device by using the on/off or ramp
key. This value is adjustable between 3 cmH
cmH
Positive Expiratory Pressure EPAP level set of your
5.0
device.
cmH
O
2
You can increase or decrease the positive expiratory
pressure level of the device by using the on/off or ramp
key. This value is adjustable between 3 cmH
IPAP value or 20 cmH
15
Ramp time.
min
You can decrease or increase the value displayed up to 30
minutes by using the ramp (down) or on/off (up) key.
Starting ramp pressure (pressure level at the start of
3.0
ramp); this parameter is adjustable and visible only if the
cmH
O
2
ramp time value is greater than 0.
The value displayed can be decreased or increased,
between 3 cmH
ramp (down) or on/off (up) key.
For further details, refer to the chapter titled "Using the
ramp" in the patient manual.
I/E ratio (ratio between Inspiratory time / Expiratory
time); this parameter can be displayed only if the Back-Up
rate is greater than 0.
This function is accessible only on the GoodKnight 425
ST device.
The value displayed can be increased or decreased by
using the on/off (up) or ramp (down) key.
However, certain conditions of Back-Up rate and
pressures can limit the possible extent of this parameter
adjustment.
Inspiratory sensitivity.
The value displayed, adjustable between 1 and 10 (1 is the
more sensitive setting), can be increased or decreased by
using the on/off (up) or ramp (down) key.
O.
2
O at the maximum.
2
O and the EPAP value, by using the
2
NOTE
O and 25
2
O and the
2
EN – 7