CAUTION
If, when in use, the device stops and displays the message Er 1, this means a problem was detected
in the settings stored. In this case, press the
« dEF » blinking message, then press the
to the default values and you must set the desired parameters again.
To exit the parameter setting menu, press again the information access button.
EN – 8
Expiratory sensitivity.
The value displayed, adjustable between 1 and 10 (1 is the
more sensitive setting), can be increased or decreased by
using the on/off (up) or ramp (down) key.
Pressure rise time (period of time over which the
pressure increases from the EPAP level to the IPAP level).
It can be decreased or increased, between 1 (about 300 ms)
and 5 (about 600 ms), by using the ramp (down) or on/off
(up) key.
However, certain conditions of Back-Up rate, pressures
and I/E ratio can limit the possible extent of this parameter
adjustment.
Hour meter value (device operating time). The value
325
displayed on the right below represents the minutes.
h
15
The hour meter value cannot be changed.
Use meter value (time after the ramp time or 15 minutes
396
h
during which the patient has breathed with the mask). The
32
value displayed on the right below represents the minutes.
The use meter value cannot be changed.
Offset of the pressure sensor integrated into the
GoodKnight 425 device. You can increase or decrease the
value displayed by using the on/off (up) or ramp (down)
key. This value can be adjustable between 0 and 255. To
set it, refer to the section "Adjustment of the pressure at
the device outlet" on page 9.
Software version.
*.*
The version number is shown in two digits (*), for
example the screen may display : Id 1.0 and the value
xx
displayed on the right below represents the motor driver
version.
NOTE
NOTE
information access button, the device displays the
on/off button The machine is then automatically reset
GoodKnight 425 / GoodKnight 425 ST