9. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under
water, as this can cause liquid to enter and damage the
device and cuff.
• If you store the device and cuff, do not place heavy ob-
jects on the device and cuff. Remove the batteries. The
cuff line should not be bent sharply.

10. Technical specifications

Model No.
BM 27
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the upper arm
Measurement
Cuff pressure 0-300 mmHg,
range
Systolic 50-280 mmHg,
Diastolic 30-200 mmHg,
Pulse 40-199 beats/minute
Display accuracy
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5% of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory
4 x 30 memory spaces
Dimensions
L 112 mm x W 110 mm x H 58 mm
Weight
Approximately 219 g (without batteries
and cuff)
Cuff size
22 to 42 cm
Permissible opera-
+10°C to +40°C, 15 to 85% relative
ting conditions
humidity (non-condensing)
Permissible sto-
-20°C to +55°C, 10-90% relative humi-
rage conditions
dity, 800-1050 hPa ambient pressure
Power supply
4 x 1.5 V
Battery life
For approx. 300 measurements, depen-
ding on levels of blood pressure and
pump pressure
Accessories
Instructions for use, 4 x 1.5V AA batte-
ries, storage bag
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
25
AA batteries