See illustration on p 50
a
Air core
Lenght
Weight
b
Latch device
c
Cable: 2 m lenght
d
Box: dimensions 140 x 64 x 28 mm - weight 200 g
e
Sliding switch with two positions (single range) or three positions (dual range)
f
Green «ON» LED
- Continuous display when battery charge is sufficient
- Flashes when under eight hours of charge remain
g
Red «OL» LED
- Displays when the measured current overloads the integrator circuit
h
Output: two Ø 4 mm male safety plugs - 19 mm between centres
Do not attach the probe to bare conductors carrying a dangerously high voltage unless wearing
protective clothing suitable for high-voltage work and using appropriate equipment for personal
protection.
The probe is not for use on cables or conductors with a potential of over 1000 V relative to earth.
Illumination of the red LED indicates an overload of the integrator circuit.
If the green LED fails to light, or flashes, the battery requires replacement. It begins
to flash when around 8 hours of battery life remain.
„ Open the flexible air-core, and attach it to the conductor to be measured.
„ Close the air-core using the latching system.
„ Centring the conductor and making the air-core as circular as possible will ensure optimum
measurement accuracy.
„ Connect the box to a multimeter with V AC or mV AC ranges and with an impedance of over 1 MW.
„ Set the switch on the box to the desired range.
„ Read off the measurement by applying the ratio indicated by the switch position on the box
(see Chapter 4.4).

2. DESCRIPTION

45 cm
< 120 g
80 cm
< 240 g

3. USE

14
120 cm
< 360 g