• Bring the lower corner up to completely cover the contents and
fold the tip back on itself to form a tab or flap (which is used later
to assist in opening the pack aseptically) (figure 5).
• Fold the left corner over the contents and fold the tip back to form a tab (figure 6).
• Fold the right corner over the left fold and fold the tip
back on itself to form a tab (figure 7).
• Bring the top corner down over the contents and tuck the corner
under the right and left folds. A small tab should be left out for
easy opening. Secure with steam indicator tape (figure 8).
Moist-Heat (Autoclave) Sterilization Process
• Pre-inspect and set up the autoclave according to manufacturer's instructions.
• Place the wrapped RPM into the steam sterilizer.
• Verify the following autoclave cycle parameters:
Method
Cycle
Temperature
Exposure Time
Drying Time
• After the sterilization cycle is complete, allow the RPM to cool before handling.
• Product is sterile and ready for use.
For detailed instructions on how to install and use the RPM with
FMX surgical instruments, refer to the Directions for Use information
packaged with each individual FMX surgical instrument.
US
Moist Heat Sterilization
Pre-vacuum
270° F (132° C)
4 minutes
20-30 minutes
EU
Moist Heat Sterilization
Pre-vacuum
273.2° F (134° C)
3 minutes
20-30 minutes