ivory Silk Line SL116C Instrucciones De Uso página 42

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

-External disinfection
Clean and disinfect the handpiece after every treatment
to eliminate residues of bodily fluids (e.g. blood, saliva
etc.) and prevent them entering the internal parts of the
handpiece.
1) Clean off all dirt and impurities from the handpiece. Do
not use metal brushes.
2) Clean the handpiece with disinfectant wipes.
3) Blow the handpiece dry with a jet of compressed air.
40
Eng
- Lubrication
CAUTION!
Lubrication cycles:
- After every 15 minutes of rotation.
- Before every sterilisation.
CAUTION!
Lubricate only with Ivory oil OL101S.
1) Fit the universal Ivory turbine handpiece adapter to
the oil spray can. Insert the other end of the universal
Ivory adapter into the chuck in the handpiece. Hold the
handpiece in place. Spray for about 5 seconds.
2) Fit the adapter for Ivory heads and contra-angles to

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Silk line sl132cSilk line sl164cSilk line sl1128c

Tabla de contenido