6.2.6

Using Stored Processing Programmes

In Manual mode: -
1.
Press and release the PROG key to select the desired programme. Either the P1, P2
or P3 icon should now be present on the display.
2.
Load the Stomacher
3.
Pressing the START key will initiate the stomaching process as previously
described.
4.
After processing, simply open the door and remove the Stomacher
instrument is ready for processing the next sample.
In Auto mode: -
1.
Press and release the PROG key to select the desired programme. Either the P1, P2
or P3 icon should now be present on the display.
2.
Press the AUTO key, displaying the AUTO icon.
3.
Load the Stomacher
4.
Closing the door will initiate the stomaching process as previously described.
5.
When the processing is completed simply open the door, remove the processed
sample and place the next sample into the instrument. Close the door and
processing will commence automatically.
6.2.7

Stall Condition

The drive controller has a factory-set electrical current limiting device to protect the motor in the
event of a stall condition occurring. In the event of a stall condition the STALL icon will be
displayed on the control panel and the instrument will stop. This condition may simply occur due to
overloading the paddles in which case the sample size should be reduced. (See 6.2.2 for LCD
icons).
Page 18 of 33
®
STOMACHER
400 CIRCULATOR
USER INSTRUCTIONS
®
Bag(s) as previously described.
®
Bag(s) as previously described.
®
Bag(s). The