FLAEM RF6 Basic2 Manual De Instrucciones página 11

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 28
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant manufac-
turer.
method B: Disinfect the accessories 1-2-8-10-11-13.1-13.2-13.3-13.4 by boiling them in
water for 10 minutes; use demineralised or distilled water to prevent calcium deposits.
method C: Disinfect the accessories 1-2-8-10-11-13.1-13.2-13.3-13.4 with a hot steam
steriliser for baby-bottle (not the microwave type). Perform the process faithfully follow-
ing the instructions of the steriliser. To ensure that the disinfection is effective, choose a
steriliser with an operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on a pa-
per towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
CLEANING IN A CLINICAL OR HOSPITAL SETTING - DISINFECTION AND STERILISATION
Before disinfection or sterilisation, sanitise the nebuliser cup and the accessories, choosing
one of the methods described below.
method A: Sanitise accessories 1-2-3-4-5-6-7-8-10-11-13.1-13.2-13.3-13.4 under pota-
ble hot water (approximately 40°C) with a gentle, non abrasive dish detergent.
method B: Sanitise accessories 1-2-3-4-5-6-7-8-10-11-13.1-13.2-13.3-13.4 in the dish-
washer with a hot cycle.
DISINFECTION
Accessories 1-2-3-4-5-6-7-8-10-11-13.1-13.2-13.3-13.4 can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium hypochlo-
rite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant manufac-
turer.
If you want to also perform the STERILISATION, jump to the STERILISATION paragraph.
After having disinfected the accessories, shake them vigorously and place them on a pa-
per towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
STERILISATION
Accessories 1-2-8-11-13.1-13.2-13.3-13.4 can be sterilised.
Device: Fractionated vacuum overpressure steam steriliser in accordance with EN 13060.
Implementation: Wrap every single part to be treated with a sterile barrier system or
packaging in accordance with Norm EN 11607. Place the packed components in the steam
steriliser. Run the sterilisation cycle according to the operating instructions of the device by
selecting a temperature of 134 °C and a time of 10 minutes first.
Storage: Store the sterilised parts as per the instructions for use of either the sterile barrier
system or packaging.
The sterilisation procedure is validated in its conformity to ISO 17665-1.
For the "CLEANING SANITISATION DISINFECTION STERILISATION" of accessories 9 fol-
low the instructions in the instruction manual.
© Copyright Gebrauchs.info
9

Publicidad

Tabla de contenido
loading

Tabla de contenido