TECHNICAL SPECIFICATIONS

RF6 Basic
Nebulizer
2
In compliance with 93/42/EEC Directive
Minimum fill volume: 2 ml
Maximum fill volume: 8 ml
APPLIED PARTS
The applied parts of type BF are:
patient accessories (1, 2, 3, 4, 5, 6, 7, 8, 9)
Ope-
FLAEM
rating
compres-
pressure
sor range
(bar)
F400
0.60
F700 -
0.80
F1000
Data shown is in accordance with Flaem Nuova internal procedure I29-P07.5.
(1)
carried out at the Inamed Research GmbH Co. KG on behalf of TÜV Rheinland LGA Products GmbH – Germany in
compliance with the new European Standard for aerosol therapy units, Standard EN 13544-1. Further details are
available upon request.
© Copyright Gebrauchs.info
MMAD
(μm)
(2)
high
standard
speed
mode
mode
3.8
4.00
3.6
3.6
10
Operating conditions:
Temperature: min 10°C; max 40°C
Air humidity: min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106KPa
Storage conditions:
Temperature: min -25°C; max 70°C
Air humidity: min 10%; max 95%
Atmospheric pressure: min 69KPa; max 106KPa
BREATHABLE FRACTION
< 5 μm (FPF) %
(2)
standard
high speed
mode
mode
62
63
66
67
In vitro characterization
(2)
DELIVERY ml/min
(1)
(approx)
standard
high speed
mode
mode
0.25
0.32
0.32
0.42