Electromagnetic Compatibility (EMC, IEC 60601-1-2, Table 2)
Electromagnetic Immunity
Calypso is intended for use in the electromagnetic environment specified below.
The customer or the user of Calypso should assure that it is used in such environment.
Immunity Tests
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power frequency
magnetic field
IEC 61000-4-8
NOTE
UT is the a.c. mains voltage prior to application of the test level.
68
IEC 60601 test level
± 6kV contact
± 8kV air
± 2kV
for power supply lines
± 1kV
for input / output lines
± 1kV
line(s) to line(s)
± 2kV
line(s) to earth
< 5% U
T
(> 95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
< 5% U
T
(> 95% dip in U
)
T
for 5 sec
3A/m
Compliance level
± 6kV contact
± 8kV air
± 2kV
for power supply lines
n/a
± 1kV
line(s) to line(s)
n/a
< 5% U
T
(> 95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
< 5% U
T
(> 95% dip in U
)
T
for 5 sec
>30A/m
Electromagnetic environment
- guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
If the user of Calypso requires
continued operation during
power mains interruptions, it is
recommended that Calypso be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic
fields should be at levels of a
typical commercial or hospital
environment.