4 Handling
4.1 Adjustments and final assembly
Static corrections are possible at any time during alignment and trial walking
as well as after completion of the prosthesis using the four 506G3 Adjustment
Screws. To exchange a module or during disassembly the previous position
may be retained by loosening only the two most deeply screwed in, adjacent
set screws.
Replace set screws that appear too long (to prevent foam cover damage) or
too short (stability) with the appropriate screw length.
Adjustment screws are available in the following sizes.
Article No.
506G3=M8x10
506G3=M8x12-V
506G3=M8x14
506G3=M8x16
Use 710D4 Torque Wrench to tighten the adjustment screws.
For trial fitting: Tighten the 506G3 Adjustment Screws clockwise
to 10 Nm.
To finish the prosthesis: Secure the set screws with 636K13
Loctite and pre-tighten to 10 Nm. Then complete assembly by
tightening to 15 Nm.
4R37
4R95
36 mm
33 mm
INOX*
Aluminum
125 kg
100 kg
–
–
Length
10 mm
12 mm
14 mm
16 mm
4R55
4R22
33 mm
33 mm
Titanium
INOX*
150 kg
125 kg
–
4R51
36 mm
Titanium
150 kg
–
–
Ottobock | 9