S With statically undefined supports, non‐uniform load
ing of the modules occurs due to flexibility as the Ioad
is taken up or due to inadequate accuracy in the
height adjustment. This non‐uniformity should be
checked an the individual Ioad cells by applying an
excitation voltage and comparing the Output voltages.
To prevent overloads, large non‐uniformities should
be compensated by inserting shims under those sup
ports which are least loaded.
S The optional cable with an outer braided wire covering
for the configurable K-Z16A3 with option 20R, is
intended for use when there is increased mechanical
stress (e.g. damage caused by gnawing rodents). If
this cable is used, the outer braided wire has to be
connected to potential equalization at least once, to
avoid accidental energization. This outer braided wire
is not used to shield the load cell. The inner braid of
the load cell cable is used for shielding.