4. Select a calibration or calibration check to show the details.
5. To show the calibrations and calibration checks for all the sensors, push
Sensors.
Change the current user
When the user of the instrument changes, change the current user setting.
Note: The user-specific settings of the instrument change when the current user setting is changed, such as the
active password and the security settings.
1. Push
until three or four dots show at the top of the screen.
2. Push the RIGHT arrow until "Surveyor HL" shows at the top of the screen.
3. Select Settings>Users.
4. Select the user, then push OK.
5. Push
, then select Set as Current User.
Identify the instrument status
Before deployment, identify the status of the instrument and the sonde. If a battery level or sensor
alert shows, complete the actions necessary to remove the alert.
1. Push
until three or four dots show at the top of the screen.
2. Push the RIGHT arrow until "Surveyor HL" shows at the top of the display.
3. Select Overview. The information that follows shows:
• Date and time setting
• Barometric pressure measured by the instrument
• Temperature measured by the instrument
• Instrument battery level
• Firmware version on the instrument
4. Push
.
5. Push the RIGHT arrow until "Sonde HL" shows.
6. Select Overview. The information that follows shows:
• Sonde battery level
• Status of the last log
• Status of the sensors
7. If a sensor alert shows, push
Monitoring
Monitoring is used for spot measuring with a deployment cable and the instrument. All
measurements are done at the same time and show in real time. Real-time measurements can be
manually captured and saved to log files on the instrument.
Real-time monitoring
Look at real-time measurements
Real-time measurements are shown in three different views on the display: list view, graph view and
grid view. Refer to
Figure
1. Push
until three or four dots show at the top of the screen.
2. Push the RIGHT arrow until "Sonde HL" shows at the top of the display.
3. Select Monitoring>Real-time. The real-time measurements show.
4. To scroll through the three different views, push the LEFT or RIGHT arrow. Refer to
16 English
and select Sensors>[Select Sensor]>Alerts to show the details.
8.
and select Show All
Figure
8.