9. Set the required sterilization cycle (Cycle 121°C or Cycle 134°C) pressing the button
SELECT.
10. Close and block the door and at the same time let the cycle start. If this is not happening
in the stabilized time the autoclave is showing the ERROR 450.
ATTENTION:
Let the cycle start up and close the door at the same time
11. Start the sterilization cycle up pressing the
The autoclave is starting the preheating phase (on the display you can read HEATING).
When the machine is reaching the necessary pressure for the chosen cycle, the
sterilization phase is starting (on the display you can read "STERILIZ").
On the display is shown the remaining sterilization time.
When the sterilization cycle is completed, the machine is discharging the pressure and
the drying phase is starting (on the display you can read "DRYING").
On the display is shown the remaining drying time.
When the sterilization cycle is completed on the display you can read "END".
Before opening the door be sure that the chamber is not under
pressure. Do not open the door if the pressure value shown on the
IMPORTANT:
display is higher than 0,03 bar
12. Open the door
1A
The closing bolt is complete with a notch to not allow the door opening
if the chamber is under pressure. To make the opening easier push the
REMARK 1
door toward the chamber;
If you do not open the sterilization chamber immediately after the end
of the sterilization cycle the steam is condensing on the door, so drops
REMARK 2
will drop while opening the door.
To avoid burnings keep at the right distance while opening the door
IMPORTANT:
and keep safe from steam losses
13. Keep the instrument from the sterilization chamber using protection gloves the special
gripper.
after that you have checked that the chamber is not under pressure.
START/STOP
button.
ENGLISH
13