8. CALIBRATION

The calibration of each Biohit Proline XL has been factory checked and
certified at 22°C using distilled water according to DIN 12650-7.
Specifications of the XL are guaranteed only when it is used with genuine
Biohit Proline tips. Due to the varying nature of different manufacturer's
serological pipettes, no specifications for serological pipettes, others than
25 ml are given. The ranges of variation of serological pipettes are listed in
the Specifications section (12).
The Biohit Proline XL can be re-calibrated in P-mode on volume ranges
5, 10, and 25 ml. Re-calibration cannot be performed on volume ranges 1
and 2 ml. The re-calibration can be performed in one selected volume at
a time in 0.05 ml increments as follows.
1. Press
2. Confirm the mode change by pressing
3. Select the desired pipetting volume to be calibrated by using
increase and
4. Press
5. Press
calibrated.
Note: When pressing both buttons the display will be blank except for
the "ml" sign. Releasing the buttons will display tthe volume to be calibrated.
6. Use
to increase or
the selected volume (e.g. for 25.00 ml volume in display program the
pipettor to give out 25.05 ml or 24.55 ml).
7. Press
8. Display will show
to give out the new volume).
9. Start pipetting.
Note: Changing the volume or mode will reset the calibration to default
settings!
to display
.
to decrease.
to confirm the selection.
and
simultaneously to display the volume to be
to confirm the selection.
and the original volume (which is now calibrated
.
to decrease the steps (in 0.05 ml steps) for
to
15