(As this procedure will empty the tip, it is recommended that you remove
the tip before resetting the XL.)
1. Clear the error message from the display by pressing
2. Place the XL in the charging base for 15 minutes.
3. Press START, which will return the piston to its home position.
Occasional "Er" situations can occur in the case that main outlets have
been switched off.They can also take place if the operating switch has
been in the OFF position during charging.
Repeated occurrence of the the "Er" message is caused by internal error,
which is due to a failure in the execution of the stroke path. In this case
the pipettor will need to be returned to the local Biohit distributor for
repair.
Problem
Droplets left inside
Leakage or pipetted
volume too small
XL out of given specs
XL blocked, aspirated
volume too small
Continuous error
message
18
Possible cause
Unsuitable pipette/tip
Pipette/tip incorrectly
attached
Unsuitable pipette/tip
Foreign particles between
adapter and pipette/tip
Silicon adapter worn out
Check the pipette/tip in use
Instrument damaged
Filter contaminated
Adapter contaminated
Instrument damaged
.
Solution
Use original Biohit tips or
pipette/tips specified by the
manufacturer
Reattach the pipette/tip
Use original Biohit tips or
pipette/tips specified by the
manufacturer
Clean the adapter
Change the adapter
Change the pipette/tip
Send for service
Change the filter
Clean or change
the adapter
Send for service