beep every second. Change the battery before further use (see Section
9.3 Replacing the Battery).
To switch OFF, press both the ON/OFF button (5) and the MODE/OFF button (6)
simultaneously.

7.2 Monitoring

The Movement Indicator Light (3) flashes with each movement during
monitoring:-
Green
Orange
Orange and
!
Movement-rate
Warning sounds
(dual-tone beep
every second)
)
Red and full
X
)
audible alarm sounds
Normal movements.
Very small or weak movements (eg very
shallow breathing).
Movement rate below 8 movements per
minute. If the movement rate increases to
over 8 movements per minute, the Movement
Indicator Light will reset to green (normal)
and the warning will stop. To disable this
function, see Section 8.3 Movement-Rate
Warning.
No movement at all has been detected for the
selected time period (15, 18 or 20 seconds).
You can select the alarm activation period (ie
the period of no movement detection) to be
15, 18 or 20 seconds (see Section 8.2 Alarm
Activation Period).
The alarm can be deactivated by briefly
pressing the MODE/OFF button (6) or by turning
Snuza Go! off completely (press ON/OFF (5)
and MODE/OFF (6) buttons simultaneously). If
the MODE/OFF button is used to deactivate the
alarm, monitoring will continue.