Advanced operation

Change the measurement units
The measurement units can be changed individually for each channel.
1. From the main menu use the
2. Use the
or
to select Measurement units and confirm.
3. Select Channel 1 and confirm.
4. Select Parameter 1 and then Parameter 2 and select one of the
following options:
Option
Description
Parameter 1
mV, pH, ISE or Disabled
Parameter 2
EC, NaCl, TDS, Ω or Disabled
5. Select Channel 2 and confirm with
confirm.
Use a sample ID
The sample ID tag is used to associate readings with a particular sample
location. If assigned, stored data will include this ID.
1. From the main menu use the
2. Use the
or
to select Sample ID and confirm.
3. Use the
or
to select
Option
Description
Automatic A consecutive number will be automatically assigned to every
sample.
Manual
A keyboard or a barcode scanner is required to enter the
sample ID name before taking a measurement (maximum
15 characters).
Select the data output
Data can be stored or transferred to a printer or to a PC.
12 English
or
to select SYSTEM. Confirm.
. Select mV, pH or ISE and
or
to select SYSTEM. Confirm.
1. From the main menu use the
2. Use the
or
to select Data Output and confirm.
3. Use the
or
to select
Option
Description
Deactivated
Select Deactivated if no printer or PC is connected.
For Printer
Select Dot matrix printer or Thermal printer.
For Computer Select Terminal, LabCom or LabCom Easy. The LabCom
Software controls several modules, pH and conductivity
meters, automatic burettes, Samplers and so on from a
computer. The LabCom Easy software gets measurements
and calibration data from the meters
Change the date and time
The date and time can be changed from the Date / Time menu.
1. From the main menu use the
2. Use the
or
to select Date / Time and confirm.
3. Use the and to advance to the next step and use the
change a value. Confirm.
The current date and time will be shown on the display.
Adjust the display contrast
1. From the main menu use the
2. Use the
or
to select Display contrast and confirm.
3. Use the and to adjust the contrast of the display and confirm.
Adjust the temperature
The temperature measurement can be adjusted at 25 °C (77 °F) and/or
85 °C (185 °F) to increase accuracy.
or
to select SYSTEM. Confirm.
or
to select SYSTEM. Confirm.
or
or
to select SYSTEM. Confirm.
to