6. Commissioning, operation
Fitting a DIN 175301-803 angular connector
1. Loosen the screw
2. Loosen the cable gland
GB
3. Pull the angled socket
4. Via the mounting hole
case
.
5. Pass the cable with the appropriate cable outer diameter (see
"Connection diagrams") through the cable gland
sealing
and the case
6. Connect the cable ends to the connection terminals on the
terminal block
"Connection diagrams" for the pin assignment).
7. Press the terminal block
8. 8. Tighten the cable gland
the cable gland and seal are not damaged and that they are
assembled correctly in order to ensure ingress protection.
9. Place the flat, square gasket
connection pins.
10. Slide the assembled angled socket
transmitter's connection pins.
11. Using the screw
transmitter, hand-tight.
20
.
.
+
from the instrument.
, lever the terminal block
.
in accordance with the pin assignment (see
into the case
around the cable. Make sure that
over the pressure transmitter's
, screw the angled socket to the pressure
out of the
, ring
.
+
onto the pressure
WIKA operating instructions pressure transmitter, model S-11
,