8.4.15
43 Advanced Code
It is possible to enable/disable the request for a PIN code.
This will remove the PIN code request in program mode and
the code 0000 will give you access to advanced settings.
Press select to confi rm your choice.
8.4.16
44 Reset Prog. settings
You can reset to the pre-programmed factory
settings for all programs. Confi rm by pressing yes.
Note! Also the programs 13-15 will reappear empty as
from factory.
8.4.17
45 Reset Filter Timer
After a fi lter change on the cassette/s, restart the fi lter timer
with this setting. Confi rm by pressing yes to reset the fi lter
time counting.
To adjust the value in the fi lter timer, see chapter 8.4.11.
8.4.18
46 Statistics
The following information is available:
Run time
Shows accumulated working hours and minutes.
Start-ups
Shows the total number of starts of the dryer.
Σ (Total Power Consumption)
Presents the total energy consumption.
Φ (Average power consumption)
Presents the average consumption for all runs.
8.4.19
47 Perform Self test
The best fault trace software on the market has been devel-
oped with this dryer. In this test all the important input and
outputs to and from the computer can be tested. This test
will give the opportunity for a quick and accurate function
verifi cation of the different parts of the dryer.
This test procedure is only available in English. By pressing
the yes button you enter the fi rst step of the self test. To
exit the self test, press the start/stop button.
Automatic testing includes the following:
8
IRT 3-1 PcD, IRT 3-2 PcD and IRT 3-20 PcD
Test 1: Push Button Test
All buttons on the control unit are tested. The corresponding
symbols are displayed by pressing the buttons. Press enter
for approx. three seconds in order to continue to the next
step of the test program.
Test 2: Display Test
Verify that all pixels light up on the display. Press enter*
and check that all pixels go out. Press enter* to continue.
* or upper left software button
Test 3: Buzzer Test
Check that the buzzer sounds. Press enter or next to
continue.
Test 4: Cassette IR Test
The IR cassette lamps light up. Check that all the IR lamps
are lit. For safety reasons this test is limited to 10 seconds.
Press enter or next to continue.
Test 5: Ventilator Test/cassette
The ventilator in the cassette now starts. A sound from the
ventilator confi rms that it is working. If the dryer is equipped
with two cassettes, you will have to repeat test no. 4 and 5
for the other cassette. Press enter or next to continue.
Test 6: Distance Sensor Test
Direct the distance sensor towards the object at a distance
of 0,3 -1 m. Check that the distance shown on the display
matches the manually measured distance. A deviation of
±3 cm is acceptable. Press enter or next to continue.
Test 7: Temperature Guards
The computer card has a temperature measuring device that
is seen on the display.
The temperature is shown in ºC or ºF depending on the set-
tings you have made.
The computer lifetime will be shortened if temperatures are
above 70 ºC/158 ºF during operation. If this happens during
drying a new window will appear after the completed drying
cycle saying "warning High temp Pc".
Test Completed
Automatic testing is now complete. Press enter or next to
fi nish.
8.4.20
48 Program Time Scale
Possibility to change between minutes and seconds.
8.4.21
49 Proc. Alarm Temp.
Not applicable on this model.
713686 rev 7

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Irt 3-2 pcdIrt 3-10 pcdIrt 3-20 pcd