Maintenance
12.9
Readout of the datalogger
In the menu Function – Logger – Read Data Logger a bar graph display with the progression
of the measured values is displayed.
The readout channel is displayed on the right hand side and can be changed.
12.10 Communications settings
The communications settings are obtained via the menu Function – Communication
Settings. These settings may only be changed if several REGARD 2400/2410 are connected
to a DRÄGER RVP 2400 visualisation panel. Up to 5 controllers can be connected to a control
desk. The addresses 1 – 5 must be assigned via the communications settings for this purpose.
13
Maintenance
13.1
Inhibit
The "Inhibit" function of the REGARD 2400/2410 blocks all the output relays in the current
status for 20 minutes. This function prevents the triggering of alarms during maintenance.
To activate Inhibit proceed as described on page 49.
60
Dräger REGARD 2400/2410