Output relay
Relay contacts
Ambient conditions
Display tolerance
Cable connection
Dimensions (H x W x D)
Material
Protection class
Approvals
(see "Approvals" on page 229)
I/O module
DC power supply
Power consumption
Communication
Input
Output
Ambient conditions
Cable connection
Dimensions (H x W x D)
Material
Protection class
Relay module
DC power supply
Power consumption
Communication
Output relays
Relay contacts
Ambient conditions
Dräger REGARD 2400/2410
: Alarm 1 and 2
: Error
: Horn
: Floating
: 250 VAC, 2 A
: Temperature: –20 °C to 60 °C
: Humidity: 10 to 90 % r.h.
: Overvoltage protection category II
: Pollution degree I
: The deviations for the display Controller Display →Transmitter
Display are <2 %
: Screw clamps for cable 1.5 mm
: 160 mm x 195 mm x 137 mm
: Plastic
: IP54
ATEX: II (2) G according to TPS 14 ATEX 1 672 X as per
EN 60079-29-1:2007; EN 50271:2010 and EN 50104:2010
EMC: EMC Directive (2004/108/EC) and Low Voltage Directive
(2006/95/EC)
: 24 VDC ±10 %
: Approx. 3 W
: RS485
: 6 digital inputs
: 6 analogue outputs
: Temperature: –20 °C to 60 °C
: Humidity: 10 to 90 % r.h.
: Screw clamps, single-wire up to 4 mm
: 90 mm x 105 mm x 72 mm
: Plastic
: IP20
: 24 VDC ±10 %
: Approx. 6 W
: RS485
: 12 freely configurable relays
: Floating 250 VAC, 2 A
: Temperature: –20 °C to 60 °C
: Humidity: 10 to 90 % r.h.
Technical data
2
2
, fine-wire up to 2.5 mm
2
65