Descargar Imprimir esta página

Ecolab Versatronic Instrucciones De Servicio Breves página 75

Dispositivo de medición modular de varios canales
Ocultar thumbs Ver también para Versatronic:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Electrical connection
Balanced connection
(when using a double-shielded connection cable in conjunction with a separate earthing
pin in the liquid)
Slot
Option/Connection option
IN 7
pH/ORP (redox)
balanced connection
IN 8
IN 9
The balanced connection serves to reduce interference by scattering electromagnetic fields
IN 10
along the sensor cable.
A separate temperature sensor can be connected to a temperature input or a universal input
for temperature compensation.
1 = glass/metal electrode (core) (A)
2 = not occupied
4 = reference electrode (inner shield) (B)
5 = liquid potential (earthing pin) (C)
6 = shield (outer shielding) (D)
Fig. 5: Terminal connection diagram pH/ORP (redox) measurement, balanced connection
417102279 Rev. 2-08.2020
24

Publicidad

loading

Este manual también es adecuado para:

255201