The current input connections are made via an ESSAILEC
neutral without opening the circuit.
The circuit to be measured must be equipped with an ESSAILEC
Remove the cover from the socket on the installation by unscrewing the two screws;
Remove the cover from the ESSAILEC
Insert the ESSAILEC
®
Connector
To start the measurements, connect the voltage output connectors to the current terminals of the power and energy analyzer. It
automatically recognizes the ESSAILEC
For the measurement configuration and technical characteristics, refer to the user manual of the analyzer you are using with
your adapter.
plug by unscrewing the two screws;
®
plug into the socket. A polarizer prevents connection errors:
test block.
®
2. Use
plug that allows simultaneous measurement on the 3 phases and the
®
current socket, identified by the green colour code.
®
Connector protection cover
Polarization
To remove the cover of the ESSAILEC
unscrew the two screws