Installation Of Wall-Mounted Master Flux Plus - Tecno-gaz Master Flux Plus Instrucciones De Uso

Ocultar thumbs Ver también para Master Flux Plus:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 133
ENGLISH
6.2

Installation of wall-mounted MASTER FLUX PLUS

Figure 11
Figure 12
(FIGURE 5, 10)
Attach the MASTER FLUX PLUS SUPPORT to the wall (13) using 4 Ø14 mm M8 wall
dowels.
Attach the MASTER FLUX PLUS PLATE to the wall (14) using 4 Ø 8(mm) wall dowels.
Screw the pin (10) onto the base of the flowmeter box.
Slot the pin into the end of the support (15).
Lock the pin to the support using the relative threaded knob.
Screw the nut on the N2O tube onto the N2O flowmeter box (H)supply nipple (RIGHT
threading).
Connect the nut on the O2 tube to the O2 flowmeter box supply nipple (LEFT threading).
Connect the free ends of the tubes to the wall outlets (press the couplings and turn to lock
the coupling to the outlet).
16

Publicidad

Tabla de contenido
loading