6.4
Evacuation of exhaled gases:
To safeguard the work environment and the staff, the exhaled gases must be conveyed
outside of the work environment.
Two systems can be used to discharge the gases externally:
-
Direct connection to the outside of the office
-
Connection to a surgical aspirator.
It is important to evaluate:
-
Whether the aspirator is suitable for the evacuation of the gases (N
-
The aspirator exhaust must be outdoors and ventilated.
Make sure it is compliant with regulations in force in the country of use.
6.4.1
Direct connection to the outside of the office
UNIT-MOUNTED MASTER FLUX PLUS
Screw the hoses (Figure 2 Pos 8/9) to the unit.
Remove the supplied valve.
Connect the mask circuit gas exhaust tube to the nozzle (Figure 15).
Connect the gas evacuation spiral tube (Figure 14) to the nozzle on the unit.
Connect the end of the spiral tube to the outside and make sure there is a final filter (Figure
14 Pos 34).
WALL-MOUNTED MASTER FLUX PLUS
Connect the gas exhaust tube to the mask circuit to the nozzle (Figure 14 Pos 36).
Connect the end of the spiral tube to the outside and make sure there is a final filter
(Figura 14 Pos 34).