Apex BiPPEX Pro Manual De Instrucciones página 17

Pulsioxímetro de mano

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 55
MD300M
±2kV for power
Electrostatic
supply lines
transient/burst
±1kV for input/
IEC 61000-4-4
output lines
±1kV differential
mode
Surge IEC 61000-4-5
±2kV common
mode
<5% UT
(>95%dip in UT)
for 0.5 cycles
Voltage dips, short
40% UT
interruptions and
(60%dip in UT)
voltage variations on
for 5 cycles
power supply input
70% UT
lines
(30%dip in UT)
IEC 61000-4-11
For2 5 cycles
<5% UT
(>95%dip in UT)
for 5 sec
Power
frequency(50/60Hz)
3A/m
magnetic field IEC
3A/m
61000-4-8
NOTE
UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration- electromagnetic immunity-
For EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The model MD300M Pulse Oximeter is intended for use in the electromagnetic specified
below. The customer of the user of the MD300M Pulse Oximeter should assure that it is
used in such an environment.
IEC 60601 test
Compliance
Immunity test
level
level
PULSE OXIMETER
±2kV for power supply
Mains power quality
lines
should be that of a typical
±1kV for input/output
commercial or hospital
lines
environment.
Mains power quality
±1kV differential mode
should be that of a typical
±2kV common mode
commercial or hospital
environment.
Mains power quality
<5% UT
should be that of a typical
(>95%dip in UT) for
commercial or hospital
0.5 cycles
environment. If the user
40% UT
of the model MD300M
(60%dip in UT)
Pulse Oximeter requires
for 5 cycles
continued operation during
70% UT
power main interruptions,
(30%dip in UT)
it is recommended that
For2 5 cycles
the model MD300M Pulse
<5% UT
Oximeter be powered from
(>95%dip in UT)
an uninterruptible power
for 5 sec
supply or a battery.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Electromagnetic environment-guidance
- 32 -
INSTRUCTION MANUAL
Conducted
3Vrms
RF IEC
150kHz
3V
61000-4-6
to 80MHz
Radiated
3V/m
RF IEC
80MHz
3V/m
61000-4-3
to 2.5GHz
NOTE1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base situation for radio (cellular/cordless)
telephones and land/mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the model MD300M
Pulse Oximeter is used exceeds the applicable RF compliance level above, the Pulse
Oximeter should be observed to verify normal operation. If abnormal performance is
observed, the additional measures may be necessary, such as reorienting or relocating the
model MD300M Pulse Oximeter.
b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
MD300M
Portable and mobile RF
communications equipment should be
used no closer to any part of the model
MD300M Pulse Oximeter, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
80MHz to 800MHz
800MHz to 2.5GHz
Where P is the maximum output
power rating of the transmitter
in Watts (W) according to the
transmitter manufacture and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,a should
be less than the compliance level in
each frequency range.b
Interference may occur in the vicinity
of equipment marked with the following
symbol:
- 33 -