The lid of the battery compartment is located on
the rear side of the appliance. Replace the three
C-type batteries by pushing the lid of the bat-
tery compartment to the rear and replacing the
spent batteries with new ones. Then carefully
close the battery compartment again.
06. Care and maintenance
Clean the device if necessary with a moist, soft, lint-free cloth.
Take care that no moisture enters the housing. Do not use
sprays, solvents, cleaning agents containing alcohol or scouring
agents; use only clean water to moisten the cloth.
Electronic equipment may not be disposed of as
domestic waste, but must be correctly disposed
of in accordance with the EUROPEAN PARLIAMENT
AND EUROPEAN COUNCIL Directive 2002/96/EC
of 27 January 2003 regarding used electrical and
electronic equipment. Please dispose of this equipment in ac-
cordance with the applicable legal regulations when its service
life is ended.
07. Technical data
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 ppm methane
Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . visual and acoustic
Warm-up time . . . . . . . . . . . . . . . . . . . . . . . . approx. 60 seconds
Response time . . . . . . . . . . . . . . . . . . . . < 22 sec. up to 40% LEL
Power supply . . . . . . . . . . . . . 3 x C-type 1.5 V alkaline batteries
Battery lifetime . . . . . . . . . . . . . . approx. 8 hours continuous use
Operating temperature . . . . . . . . . . . . . . . 0 to 50°C (32 to 122°F)
Storage temperature . . . . . . . . . . . . . . -30 to 60°C (-22 to 140°F)
Operating humidity . . . . . . . . . . . . 10 - 90% RH (non-condensing)
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221 x 72 x 46 mm
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 520g
Moving the appliance from a cold to a warm environment (and vice
versa) can lead to the formation of condensation on the measuring
circuitry of the appliance. This physical effect, which cannot be pre-
vented by structural means for any measuring instrument, leads to
false measurement values. Depending on the extent of the tempe-
rature difference, the appliance will require an 'acclimatisation time'
of approx. 15-30 min. before the measurement procedure can be
resumed.
m
Extreme humidity or heat can damage the sensor.
Operating Manual – Gas Detector BG 30
min. 10% UEG/LEL, max. 40% for methane