ALLURIS FMT-100WT Instrucciones De Servicio página 21

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Calibration
For recalibration, you would need a solid and vibration-free rack and a test weight (50kg), preferably Class M1,
which complies with the international standard.
Switch off the instrument using ON/OFF-button.
Press and hold MODE- PEAK- and RESET-buttons
simultaneously.
While keeping the above buttons pressed, press the
ON/OFF-button until CAL is indicated in the display
and the calibration weight code (see table) is shown.
Then release all buttons.
Press the MODE-button. The signs SNC and –00-
flash for approx. 10 seconds followed by the
indication ZER 88888.
To confirm the calibration of the zero point press the
RESET-button. The indication SCN 88888 flashes on
the display for approx. 15 seconds, followed by PEK
88888. The first step of the calibration is now
complete.
Hang the calibration weight on the terminal adaptor.
When the weight hangs perpendicular and still, you
may save the full-scale value.
Press the Peak-button. The display flashes for
approx. 15 sec. with the sign SCN 88888 followed by
the sign END 88888. After a further period of approx.
5 sec., the sign OK 88888 indicates that the
calibration was successful. The instrument switches
off automatically.
Should the calibration fail, the display will show ERR
88888 instead. The original calibration datas are still
saved in the memory, allowing you to repeat the
calibration procedure.
On/Off
Reset | Mode | Peak
On/Off
Mode
Reset
Peak
21
CAL
-00-
ZER
88888
PEK
88888
END
88888
OK
88888

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Fmt-110wt

Tabla de contenido