ALLURIS FMT-100WT Instrucciones De Servicio página 22

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Technical Data / Maintenance
Measuring Range
Wire Diameter
Resolution
Accuracy
Operation Mode
Overload
Display
Memory
Power Supply
Interface
Temperature
Range
Protection Code
Weight
Dimensions
Housing Material
This instrument is maintenance-free under proper use. Torn parts of wire must not fall into the spaces around
the connection adapter; should this occur, blow them off or remove them using the vacuum cleaner.
Terminal adapter width
Clamping device
Dia. range IEC 60352-2
Dia. range SAE AS7928 II
@ 23°C (F.S.)
Tk (absolute)
Tk (relative)
Standard
Peak
max. admissible
Type
Update time (Standard)
Update time (Peak)
Type
RS232C
Operation
Storage
LxWxH (without lever)
0,0...500,0 N
0,5 | 0,8 | 1,0 | 1,3 | 1,5 | 2,0 | 2,5 | 3,0 | 3,5 | 4,0 | 5,0 | 6,0 mm
0,1 ... 6,0 mm (continuously variable)
cross section 0,05 ... 10 mm² (AWG8...30)
AWG 12 ... 28
0,1 N
+/- 0,5% (+/- ½ digit)
automatic readjustment while starting
+/- 0,02% (°K)
displays actual value in N | kgf | lbf
displays peak value in N | kgf | lbf
200 % (F.S.) | alarm at 120% (F.S.)
LCD, 4-digit, 12mm high
1000 msec | 500 msec | 333 msec | 200 msec
100 msec | 50 msec (selectable)
1 msec
peak value (Peak)
internal NiCd-battery (up to 12h)
AC adapter-/charger 100...240VDC (50...60Hz)
Baud rate selectable 2,4 kB | 4,8 kB | 9,6 kB | 19,2 kB
0°... 40° C
-20°... 60° C (rF < 80%)
IP 40
approx. 14 kg
350 x 160 x 75 mm
anodised aluminium, surface tempered steel,
stainless steel V2A
22

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Fmt-110wt

Tabla de contenido