nm - nanometer: A unit of length (one billionth
of a meter) commonly used to describe the
wavelength of visible light energy. The Optilux
Radiometer uses filters that only allow light
energy between 400-500 nm to be measured.
400 - 500 nm is the violet and blue portion of
the visible spectrum.
2
mW/cm
- milliWatt per square centimeter:
A unit of power density used to describe the
concentration of energy distributed over a
2
particular area.
mW/cm
is the generally
accepted standard unit for specifing the output
of a dental curing light unit. The meter scale on
the front of the Optilux Radiometer is calibrated
2
from 0 -1000 mW/cm
(Figure 2). The higher
the reading, the more blue light output.
4.
SPECIFICATIONS AND
EQUIPMENT
CLASS
a)
Equipment class:
I
b)
Protection from
None
ingress of liquids:
Environment:
Equipment not suitable
c)
for use in the presence
of
a flammable anesthetic
mixture with air or nitrous
oxide
d)
Transport and storage:
• Ambient temperature
-40° F to 158°F
(-40°C to +70°C)
• Relative humidity
10% to 100%,
condensation
• Atmospheric
.5 atm to 1.0 atm (500
pressure
hPa to 1060 hPa)
4
e)
Operation:
f)
Sensitivity range:
(nanometers)
g)
Intensity range:
h)
Accuracy:
i)
Unit overall dimensions
• Height
• Length
• Width
j)
Unit weight:
k)
Additional symbols:
Attention, consult
accompanying
documents
5.

NEED FOR RADIOMETER

In
normal
use
deteriorate from a number of causes, even
before the lamp fails, such as:
• Loss of reflectivity of the lamp reflector
(Figure 3).
• Blackening or frosting of the lamp filament
envelope (Figure 4).
• Contamination
coatings (Figure 5).
• Cracked or missing filters.
• High
AC
outlet
conditions.
including
• Very high output lamps.
Continuous operation
400-500 nm
2
0-1000 mW/cm
±5% of reading
1.25"
(3.175 cm)
4.75"
(12.06 cm)
3.00"
(7.62 cm)
4.5 oz. (127.5 g.)
curing
light
output
can
or
deterioration
of
filter
line
voltage
(mains)
5