EN-12
Source = ABS - Displayed Weight, Output Type = 4-20 mA
Under Load
Over Load
Source = Gross Weight, Output Type = 4-20 mA
Under Load
Over Load
Source = Displayed Weight, Output Type = 0-10 VDC
Under Load
Over Load
Source = ABS - Displayed Weight, Output Type = 0-10 VDC
Under Load
When the ABS - displayed weight increases, the analog output
signal keeps increasing towards the maximum output (20 mA).
When it reaches the under load point, the analog output
switches to approximately 0 mA immediately. In this kind of
situation, the output remains 0 mA, or it will output correctly.
It is possible only when weighing is in the gross mode; it
works the same way as the standard "Displayed Weight"
setting in this case.
When the gross weight drops below zero, the analog signal
keeps decreasing. When it reaches the under load point, or it
is lower than the lower limit of analog signal range, the analog
output will switch to approximately 0 mA immediately. In this
case, the output remains 0 mA, or it will output correctly.
When the gross weight exceeds the over load, the analog
signal continues to increase. When the over load point is
reached, or the analog signal positive range is exceeded, the
analog output immediately switches to approximately 24 mA
and remains there until the display weight is no longer over
load or the analog signal returns to within the range.
When the displayed weight (gross or net) drops below zero,
the analog signal keeps decreasing. When it reaches the
under load point, or it is lower than the lower limit of analog
signal range, the analog output will switch to approximately -
2.4 VDC immediately. In this kind of situation, the output
remains -2.4 VDC mA, or it will output correctly.
When the displayed weight (gross or net) exceeds the over
load,
analog signal keeps increasing. When it reaches the over load
point, or it is higher than the higher limit of analog signal range,
the analog output will switch to approximately 12.5 VDC
immediately. In this kind of situation, the output remains 12.5
VDC mA, or it will output correctly.
When the ABS - displayed weight increases, the analog output
signal keeps increasing towards the maximum output (10
VDC). When it reaches the under load point, the analog output
switches to approximately -2.4 VDC immediately. In this kind
of situation, the output remains -2.4 VDC, or it will output
correctly.
the