6. Care and Maintenance
6.2 Refilling the Water Softening Salt Reservoir

6.3 Changing the Air Filter

When the message "Replace air filter" appears, the Hepa filter is in need of changing. Turn
the power off. Remove the four screws from the filter door on the right hand side of the
unit. Remove the old filter by pulling the left side of the filter out first, and then removing it.
Install the new air filter and replace the filter door.
After replacing the filter, go into the "Reset Drying Counter" screen in the setup menu
(section 5.9) and reset the drying counter to zero.
The filter needs to be replaced every 750 drying cycles. The unit will continue to run if the
filter is not replaced, but the Hydrim's drying performance will decrease.
Page 21
Check the salt window on the front panel periodically.
When the level is low, refill the salt reservoir.
To fill the salt reservoir, follow these steps:
1. Open the salt container lid.
2. Fill the salt container to the top
(maximum 0.5 kg)
3. Close the salt container lid.
4. Run a complete cycle, with no instruments.
HYDR IM C51wd Operator's Manual