10. Specifications
Height 475 mm
Width: 600 mm
Depth: 460 mm
Depth with door open: 780 mm
Weight: 35 kg
Running Noise: 60 dB(A)
Hot and cold water connections: G 3/4"
Inlet water pressure: 1-10 bar
Incoming hot water temperature: 60°- 84°C
Drain: 3/4"
Water softener: 0.5 kg salt capacity
Drying System: Heater 1 kW
Electrical Connection: 220-240 V 50 Hz 10 A
Equipment pollution degree: Pollution Degree 2
Equipment installation category: Installation category II
Maximum relative humidity: 80% for temp up to 31°C
50% for temp up to 40°C
Operating temperature range: -5°C to 40°C
Max. altitude: 2000m
Mains supply: + / -10% of nominal
Required clearance for ventilation: 75mm on either side, at the top and at the
rear of the unit.