SciCan HYDRIM C51wd Manual Del Operador página 30

Ocultar thumbs Ver también para HYDRIM C51wd:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

7. Troubleshooting
Error Messages
Error
CF1 Water Heating
failure
CF2 Chamber Filling
Failure
CF3 Chamber temper-
ature reading failure
CF4 Water evacuation
failure
CF5 Disinfection
Failure
CF6 Serial
Communication
Failure
CF8 Secondary
Chamber Temperature
Sensor Failure
CF9 Software or PCB
Failure
CF13 Temperature
Validation Error
CF15 Chamber
overflow
HYDR IM C51wd Operator's Manual
Cause
The water does not reach the
required temperature in the
specified time
The water does not fill the
chamber within the specified time.
Faulty sensor.
Water cannot drain from the unit.
Temperature dropped below 80
during the disinfection phase.
Failure the read the second
temperature sensor within 10 sec.
Second temperature sensor
readings are out of limits.
Cycle length is more than 2hr 30
min.
Second temperature sensor
readings are out of limits.
Overflow switch did not turn off after
30 sec of running the drain pump.
Solution
Call for service.
Check that the water hoses
are not kinked. Open the
water shut-off valves.
Call for service.
Check that the drain hose is
not kinked. Check that the
filters in the bottom of the
chamber are not obstructed.
Call for service.
C
˚
Call for service.
Call for service.
Call for service.
Call for service.
Check if drain is blocked;
Call for service.
Page 24

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido